Precision and recovery describe two features of a method
Eurachem and CITAC’s 2026 guide explains how laboratories can use their own validation measurements to estimate uncertainty. Precision describes the spread of repeated results under defined conditions. Recovery asks how well the method accounts for a known reference amount in a relevant material. Results can cluster tightly even when the method consistently misses some of what it is meant to measure.
The matrix is the rest of the material around the substance being measured. Depending on the test, differences in composition can change how much reaches the detector or how strongly it registers.
A second sample can change the answer
For a simplified illustration, suppose a measurement process returns 90 of 100 reference units in one sample type, and 60 of 100 in another. A correction inferred from the first sample would leave a substantial error in the second. The arithmetic may be internally consistent while the assumption about the sample fails.
The guide discusses using multiple relevant reference materials to detect differences in recovery. It accounts for uncertainty in the recovery estimates themselves, so differences should be interpreted alongside their precision and the suitability of the reference materials.
Report the range the evidence covers
Validation records help users see the sample types and concentration range for which a method’s uncertainty applies. The guide discusses incorporating matrix-specific recovery or uncertainty when conditions warrant it. A corrected number still carries uncertainty from both the measurement and the estimate used to correct it.